VNG-RW-900

Gallery:-

FEATURES:-

Heater Module

Controller

Examination Light

Apgar Timer

Alarm

Probe Fail

PRODUCT DESCRIPTION:-

ELECTRICAL SPECIFICATION:

  1. Skin:32 C to 380
  2. Resolution Accuracy:+0.2
  3. Power source – 220 Volt Ac 50 Hz
  4. Heater power:650 Watt
  5. Examination light: 9 Watt
  6. Power cable length – 3 Meter
  7. Fuse used:Amp S9+ ho 

HEATER MODULE
The radiant energy is directed by the parabolic reflectors on the intant. The radiant
energy can be controlled either manually or by mean of skin temp sensor which
controls the radiant energy to maintain a selected skin temperature.

CONTROLLER
when operated in manual controt mode, the controller permits the operator to
adjust the heat output of the heater module from zero to maximum settings. During
the manual mode, a patient probe can be used to monitor the skin temperature
when operated in skin temperature contral mode, the controller utilizes askin
temperature probe connected in between the controller input and the intant to
automatically adjust the heat output of the heater module to maintain a selected
preset skin temperature. Up and down switches permit adjustment of skin temp set
point and a digitai display provices temperature read out.

EXAMINATION LIGHT
Dazzle tree observation light tor examination of the ntant
Dual side swivel heater provided access 1or x-tay and
other procedure

APGARTIMER
Apgar mode is used to take tne intant Apgar Score wiEnin
specitied time
Apgar mode is to start and stOp just pressing Apgar switch

ALARM
Each time the unit is turned on, an automatic test sequence is
initiated to verify that the visual display and the audible
alarms are functional. Alarms are provided for high and low
skin temperature.

SET POINT (HIGH)
If the temp sensed by the skin probe is a 5 degree above the set temperature, ahigh
set point alarm will occur after a delay of 10 second. The audible portion of the alarm
can be silenced by depressing the alarm mute switch. However the visual indicator
will continue. If the condition is not corrected within 20 sec the audible alarm will
resume. When the alarm condition is corrected, the circuit will reset

SET POINT L (LOW)
If the set temperature sensed by the skin probe is 0.5° less than the set point, a low
set point alarm will occur after a delay of 10 seconds.

PROBE FAIL
During skin control mode, if the skin temp probe is open short or un-plugged, the
probe failurealarm will indicate the baby temperature display and the heater will be
turned off. This alarm cannot be rest until the alarm condition is corrected

MODES OF OPERATION

Servo mode
in this mode infant temperature is sensed with the help of a skin temperature probe.
The heater power is automatically controlled so that intant temperature lies in the
desired range. This is set by the user.

Manual mode
in this mode, we can manually set the power of the heater ranging
from 10 percent to 100 percent