Double Surface Phototherapy

Technical Specification of Phototherpy Unit Double Surface
Sr.No. Specification
1 LED lamp should have Minimum 9 hi power LEDs (Blue) (3 white LED for observation) providing the same intensity level
2 Irradiance should be >40 microwatt/cm2 /nm at recommended distance.
3 Light source’s life time should be minimum 30,000hrs
4 There should be less than 10% change in illumination after 30000 hrs (irradiance).
5 Wavelength spectrum should be between 420-470nm with peak wavelength around 450nm.
6 Effective surface area should be a minimum of 50cm x 25cm at the recommended distance.
7 Irradiance ratio should be >0.4(minimum/maximum Irradiance )
8 There should be a machine run time totalizer and Infant therapy time totalizer
9 Over temperature cut off safety feature should be present
10 Height should be adjustable between 120 cms to 160 cms
11 Electrical supply : Input voltage 100-240V, 50-60Hz.
12 It should be a combination of over head Phototherapy Unit and undersurface Phototherapy with Baby Bassinet
13 Baby Bassinet should have :
a Clear collapsible fold down side panels.
b Baby Tray with transparent base for Undersurface Phototherapy usage.
c X-Ray cassette guide facility.
14 Same as over head LED box Under surface Phototherapy LED box should have a LCD Digital Timer for lamp usage hours and patient exposure
15 Unit should be Indian CE Standard/European CE/ISO approved
16 Demonstration must be done by the bidder at the time of technical evaluation meeting.