VNG Infant Incubator 3000

With an ambient temperature probe, the incubator is able to compare the gap between ambient and setting tempera- ture and changing heating power output accordingly in order to provide a more stable inside hood air temperature.

Category:
Physical AttributesHeight WidthDepth
Hood Specification
Tubing access ports Access door sizemattress to hood height 
              10             180x 13cm4.8cm
———————————————————-—————————
Soft bed mattress sizeMattress tiltDrawer depth 
73.6 x 38.6 x 1.8 cm±12°36cm
——————————–————————–————————
        Drawer size  
28.0 x 2407cm         28.0 x aa.3cm (x2)  
       
Temperature control Modes
Temperature control modes Air mode control temperature rangeAir mode control override temperature range
Skin and air temperature control mode20°C-37°C37°C-39° C
————————-———————-———————
Air mode control accuracy Skin mode control temperature range Air mode control override temperature range 
 34°C-37°C37° C- 38°C
≤±1.0°C—————————-—————————-
—————————————Dual-skin temperature monitoring  
  Skin temperature accuracy yes  
±0.3°C  
    
Servo Humidity option 
Humidity control range Humidity control operating time without refilling Humidity control reservoir capacity 
 24Hours 1500ml
30%-95% RG———————————-——————————
——————————–Humidity control accuracy  
Humidity display accuracy   
±5% RH ±5%RH 
    
Performance 
Air flow velocity across mattressTemperature rise time at 22°C (72°F) ambientTemperature variability 
<10cm/sec<20min (frpm22°C, 50%RH,  to 35°C)<0.2°C
——————————————–—————————————————– 
Temperature Overshoot Temperature uniformity with a level mattress —————————————
<0.5°C maximum 0.8°COperating noise level in hood
  ≤45dBa
————————————————————————————– 
        Carbon Dioxide (CO2)Level  ——————————
                  <0.5%  
Servo Oxygen Option 
Oxygen display accuracy of full scale Oxygen control accuracy(100% calibration)Oxygen control accuracy(21% calibration)
±2%±3%±5%
————————————–—————————————-————————————-
Oxygen control range Oxygen display resolution  
21% to 65%1%  
Scale Option 
Weight range Weight display resolution Weight accuracy 
300g to 8kg 1g±10g
Operating Environment 
Temperature Humidity Air Velocity 
20° to 30°C10 to 95%RH Up to 0.3m/sec
Storage/ Shipping Information 
Temperature Humidity  
-25°  to 60° C0 to 95% RH