VNG Neotherm ™

There is one established treatment that can minimize permanent brain damage from hypoxic-ischemic encephalopathy (HIE). However, it must be given very shortly after birth/the oxygen-depriving incident in order to be effective (ideally within six hours.) This treatment is known as hypothermia therapy, but it has many other names. Such as “therapeutic hypothermia,” “cooling therapy,” and “neonatal cooling.” Hypothermia
therapy involves cooling the baby down to a temperature below homeostasis to allow the brain to recover from a hypoxic-ischemic injury. Typically, the target temperature is about 33.5 degrees Celsius (92.3 Fahrenheit)

General Specifications

S. No.

  

1.

Dimension 

(LxHxW) 460mm x 30mm x 530mm

2.

Weight

23 Kg

3.

Electricity Consumption

1.5A/230 VAC, Max. 345 W

4.

Height With Trolley

1092mm

5.

Noise output

45- 49.5 dB (Mean value 1m distance)

6.

Fluid capacity

500ml (Inner Tank). 500ml (Fill Up Set)

7.

Fluid active ingredients

Distilled Water

8.

Minimum Fluid Temperature

12°C

9.

Maximum Fluid Temperature

39°C

10.

Maximum Patient Weight

10kg

   

Technical Specifications

S. No.

Description

1.

It is a Micro-Processor based servo – controlled neonatal cooling-warming system

2.

It is based on TEC Technology.

3.

It has single mattress for the whole body of neonates.

4.

It has a single tank with a capacity of 500 ml and be used for at least 2 to 3 patients.

5.

It has tendency to cool down body up to 30° C

6.

It has ability to re-warm body to normal temperature at a user selected rate (36.5°C).

7.

It monitors the patient via rectal temperature and use that for servo control mode.

8.

It has continuous display of Set temperature, Measured temperature of Rectum, Measured temperature of Skin, Measured temperature of Mattress.

9.

Alarms for high and low temperature if deviation from target temperature>1° C, Water flow Failure & Power Failure.

10.

The system is mounted on a sturdy compact trolley with castor wheels and brakes of elite quality.

11.

It has ability to transfer data to portable media/computers. The data is transferred in the form of .CSV File.

12.

It has a Memory of set and measured parameter for at least 72 hours.

13.

This cooling system uses distilled water for operation.

14.

The Company always mentions the make & model name of quoted equipment and submit the technical brochure of quoted the model in the technical bids along with compliance sheet as per technical specification.

    1. Rectal temp. probes: 5 Nos. with each system
  • b. Skin temp. probes: 5 Nos. with each system
  • c. Reusable cooling mattress for baby: 5 Nos. with each system

15.

Warranty for 3 years and post Warranty CAMC for 7 years

16.

The Company certifies that model quoted is latest and not obsolete and spares will be available for next 7 years after the completion of warranty.

17.

The demonstration and training to hospital staff will be given at the time of installation.